Sumber Aneka Karya Abadi - Your trusted partner for laboratory instrument

Search

MEMMERT, Climate chamber ICH110 with humidity 10%-80% rh

Prod #ICH110

CLIMATE CHAMBER Memiliki 3 tipe ukuran volume : 108 L , 256 L, 749 L ; Terdiri dari 3 model: climate chamber ICH 110/260/750 temperature/humidity , ICH 110L/260L/750L dengan cahaya lampu , ICH 110C(260C/750C dengan kontrol CO2 ; Range Temperatur : * - 10 °C s.d. + 60 °C (model ICH) * 0 °C s.d.+ 60 °C (model ICH L and ICH C) ; 4.Twin display ; 5.Kontrol Digital untuk kelembapan 10-80% rH ; 6.Double door (pintu Kaca)

Product Overview

Properties

Control of standard components

Temperature 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
Humidity humidity supply with distilled water from external tank by self-priming pump
Humidity humidity supply with distilled water from external tank by self-priming pump
Humidity humidification by hot steam generator
Humidity active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity - resolution of display 0.1 %, setting accuracy 0.5 %
ControlCOCKPIT adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days

Temperature

from -10°C up to +60°C without humidity
from +10°C up to +60°C with humidity
resolution of display for setpoint and actual temperature values 0.1°C

Control technology

Function SetpointWAIT the process time does not start until the set temperature is reached
Language setting German/English/Spanish/French
Calibration three freely selectable values each, temperature/humidity/CO2
adjustable parameters temperature (Celsius or Fahrenheit), fan speed, relative humidity, programme time, time zones, summertime/wintertime

Ventilation

Fan forced air circulation in working chamber adjustable from 10 to 100 % in 10 % steps

Communication

Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
Temperature control over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature
Autodiagnostic system integral fault diagnostics for temperature and humidity control
Alarm visual and acoustic

Heating concept

no drying-up of the load caused by the cooling device due to separation of thermal jacket from interior
energy-saving use of CFC-free cooling/heating system (refrigerant R134a)
highly efficient automatic defrosting system
air jacket heating system for gentle all-around heating

Standard equipment

Door fully insulated stainless steel door with 2-point locking (compression door lock)
Door inner glass door
Housing rear zinc-plated steel
Internals 2 stainless steel grids
Installation on lockable castors

Stainless steel interior

Max. loading of chamber: 150 kg
w(A) x h(B) x d(C): 560 x 480 x 400 mm
Volume 108 l

Textured stainless steel casing

w(D) x h(E) x d(F): 745 x 1233 x 634 mm

Electrical data

Voltage Electrical load 230 V, 50/60 Hz
approx. 1350 W
Voltage Electrical load 115 V, 50/60 Hz
approx. 1350 W

Packing/shipping data

the appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton B x H x T: 880 x 1410 x 810 mm
Net weight approx. 109 kg
Gross weight carton approx. 137 kg