Specifications :
Accuracy:
Turbidity up to 1000 NTU:
without calibration < 5% of the measured value ±0.01 NTU
with calibration < 1% of the measured value ±0.01 NTU
Automatic Wiper : Yes
Body Material : Stainless Steel
Cable Length : 10 m (optional extension cables available)
Calibration Method :
Turbidity: Formazin or STABLCAL Standard (at 800 NTU).
Suspended Solids: Sample specific, based on gravimetric TSS analysis with a correction factor procedure.
Certifications : CE
Configurations : Sensor only
Controller : Sensor only
Diameter Sensor : 60 mm
Flow : Max. 3 m/s (the presence of air bubbles affects the measurement)
Length : 200 mm
Maintenance Interval : 1 h/month
Material :
Optics Carrier and Sleeve: stainless steel 1.4571; Wiper Arm: stainless steel 1.4581; Wiper Rubber: silicone rubber (standard) Optional: Viton (LZX578); Wiper Shaft: stainless steel 1.4571
Max. cable length Sensor-to-Controller : 100 m
Measuring Principle :
Infrared Duo scattered light technique for color-independent turbidity measurement Turbidity in accordance with DIN EN ISO 7027 / TS equivalent DIN 38414
Model : Solitax ts-line sc
Mounting : Immersion
Operating Temperature Range : 0 - 40 °C
Parameter : Suspended Solids, Turbidity
Pressure Range : Stainless steel immersion sensor: <6 bar or <60 m
Range :
TSS-content: 0.001 mg/L - 50 g/L
Turbidity: 0.001 - 4000 NTU
Repeatability :
TSS content: < 3 %
Turbidity: < 1 %
Response Time T90 : 1 - 300 s adjustable
Special notes :
installation note: distance sensorwall > 50 cm (Turbidity);
distance sensorwall>10 cm (TSS),
Weight : 1.38 kg
What's included? : Turbidity/Suspended Solids probe, user manual